Thermal Cycler : C1000 Chases, Maximum ramp rate, °C/sec : 5, Average ramp rate, °C/sec : 3.3, Heating and cooling method : Peltier. Lid, °C : Heats up to 105
Temperature
Range, °C : 0-100
Accuracy, °C : ±0.2 of programmed target at 90°C. Uniformity, °C : ±0.4 well-to-well within 10 sec of arrival at 90°C
Gradient
Operational range, °C : 30-100, Programmable span, °C : 1-24.
Included :
Touch thermal cycler chassis, optical reaction module, real time manager software, license for qbase software, communication cable, ups, computer, HRM Software
PC = Komputer Dell
Printer Ink Jet
UPS 2200VA, 230V
Optical Detection
Excitation : 6 filtered LEDs ( FAM, SYBR, HEX, TET, Cal gold 540, VIC, ROX, TEXAS RED, Cal red 610, Cy5, Quansar 670, Quansar 705)
Detection : 6 filtered photodiodes
Range of excitation/emission wavelengths, nm : 450-730
Sensitivity : Detects 1 copy of target sequence in human genomic DNA
Dynamic range : 10 orders of magnitude
Scan Time
All channels, sec : 12
FAM/SYBR® Green only, sec : 3
Software
Operating system: Windows XP, Windows Vista Windows 7
Multiplex analysis : Up to 5 targets per well
System
Licensed for real-time PCR : Yes
Sample capacity, wells : 96
Sample size, μl : 1-50 (10-25 recommended)
Communication interface : USB 2.0
Electrical approvals : IEC, CE
Dimensions (W x D x H), cm/in : 33 x 46 x 36 /13 x 18 x 14
Weight, kg/lb : 21/47
Minimum PC Specifications :
Opertaing system : Windows XP Professional SP2,
Processor speed : 1 GHz
RAM : 1 GB ( 2GB for Vista)
Hard Drive Space : 10 GB
Screen Resolution : 1,024 x 768 with true-color mode